 learn how to perform a HART pressure transmitter calibration from a downloaded task with the Fluke 729 Automatic Pressure Calibrator.

1. Connect your 4-20 milliamp jacks to your Fluke 729 and the HART transmitter

2. Press the VENT button to zero out the pressure for accurate measurements

3. Press the HART button

4. Enable your 250 ohm resistor and loop power

5. Once the Fluke 729 is in sync with the HART transmitter, press continue

6. Scroll through and select appropriate task for the calibration

7. Press F4 to recall HART transmitter

8. Configure settings for calibration

9. Press F4 to begin Auto Test

The Fluke 729 Automatic Pressure Calibrator has been designed specifically with process technicians in mind to simplify the pres­sure calibration process and provide faster, more accurate test results. Technicians know that calibrating pressure can be a time-consuming task, but the 729 makes it easier than ever with an internal electric pump that provides automatic pressure generation and regulation in an in an easy-to-use, rugged, portable package.
The ideal portable pressure calibrator, the 729 allows you to simply type in a target pressure, and the calibrator will automatically pump to the desired set-point. Then, internal fine adjustment con­trol automatically stabilizes the pressure at the requested value.

The Fluke 729 can also automatically test multiple pressure test points and automatically document the results. Calibration is as easy as typing in the starting and ending pressure and the number of test points and tolerance level. The 729 does the rest.

Built-in HART communication capabilities enable HART transmitter mA adjustments, light HART configuration and the ability to adjust to applied 0% and 100% values.

Upload and manage documented calibration results with DPCTrack2™ Calibration Management Software, making it easy to manage your instrumentation, create scheduled tests and reports, and manage calibration data.

With three ranges to choose from, 30 psi (2 bar, 200 kPa), 150 psi (10 bar, 1 MPa) and 300 psi (20 bar, 2 MPa), Fluke 729 Automatic Pressure Calibrators are designed to perform when and where you need them.

Some key features include:

• Automatic pressure generation and regulation to 300 psi
• Easily document the process using onboard test templates
• Automatic internal fine-pressure adjustment
• Measure, source and simulate 4 to 20 mA signals
• 24V loop power for powering transmitters for tests
• HART communication for testing HART smart transmitters
• Built-in leak testing for verifying test setups for leakage


 learn how to perform a HART pressure transmitter calibration from a downloaded task with the Fluke 729 Automatic Pressure Calibrator.

1. Connect your 4-20 milliamp jacks to your Fluke 729 and the HART transmitter

2. Press the VENT button to zero out the pressure for accurate measurements

3. Press the HART button

4. Enable your 250 ohm resistor and loop power

5. Once the Fluke 729 is in sync with the HART transmitter, press continue

6. Scroll through and select appropriate task for the calibration

7. Press F4 to recall HART transmitter

8. Configure settings for calibration

9. Press F4 to begin Auto Test

The Fluke 729 Automatic Pressure Calibrator has been designed specifically with process technicians in mind to simplify the pres­sure calibration process and provide faster, more accurate test results. Technicians know that calibrating pressure can be a time-consuming task, but the 729 makes it easier than ever with an internal electric pump that provides automatic pressure generation and regulation in an in an easy-to-use, rugged, portable package.
The ideal portable pressure calibrator, the 729 allows you to simply type in a target pressure, and the calibrator will automatically pump to the desired set-point. Then, internal fine adjustment con­trol automatically stabilizes the pressure at the requested value.

The Fluke 729 can also automatically test multiple pressure test points and automatically document the results. Calibration is as easy as typing in the starting and ending pressure and the number of test points and tolerance level. The 729 does the rest.

Built-in HART communication capabilities enable HART transmitter mA adjustments, light HART configuration and the ability to adjust to applied 0% and 100% values.

Upload and manage documented calibration results with DPCTrack2™ Calibration Management Software, making it easy to manage your instrumentation, create scheduled tests and reports, and manage calibration data.

With three ranges to choose from, 30 psi (2 bar, 200 kPa), 150 psi (10 bar, 1 MPa) and 300 psi (20 bar, 2 MPa), Fluke 729 Automatic Pressure Calibrators are designed to perform when and where you need them.

Some key features include:

• Automatic pressure generation and regulation to 300 psi
• Easily document the process using onboard test templates
• Automatic internal fine-pressure adjustment
• Measure, source and simulate 4 to 20 mA signals
• 24V loop power for powering transmitters for tests
• HART communication for testing HART smart transmitters
• Built-in leak testing for verifying test setups for leakage